HDT1500/HDT3000: Brinell
Venligst kontakt os for pris.
HDT1500/HDT3000: Brinell
Case Hardness Depth (CHD) analyzers HTD1500 & HTD4000
HTD1500 and HTD4000 are case hardness depth analyzers, able to verify thicknesses up to 1.3mm (HTD1500) and 2.7mm (HTD4000) in an immediate and non-destructive way.
The instrument, designed to meet the requirements of heat treatment control, has a large capacity stand and is able to perform the test of pieces of large dimensions; does not require particular maintenance and has been studied to work in a particularly critical environments.
The carbide penetrator has a service life of about 2500-3000 tests, depending on the type of material and the maximum load used.
Product advantages HDT1500/HDT3000: Brinell
- Ernst patented principle. Case hardness depth determination from 0.05mm to 2.7mm.
- Non-destructive testing. Testing of the whole production batch is possible eliminating disposal costs aand allowing savings on consumables.
- 1 test: 1 minute. Drastic reduction of the test times, increased productivity.
- One measurement vs. a complex process. No need for cross-sectioning, polishing and mounting of the test piece.
Technical details HDT1500/HDT3000: Brinell
Power supply | 110-240 Vac 50/60 Hz |
Loads | MOD HTD1500: 100 kgf – 1500 kgf (N 981 – N 14715) MOD HTD4000: 200 kgf – 4000 kgf (N 1961 – N 39227) |
Norms and certifications | ASTM E-18 indirect method on certified test block indirektes ISO 6508 indirect method on certified test block indirektes |
Minimum measurable diameter | 10 mm |
Overall dimensions | 720 mm (ca. 950 mm incl. PC) x 1200 mm x P 880 mm |
Working area dimensions | Stand capacity 335 mm – H 315 mm |
Selectable functions | load selection – language – calibration – hardness measure – schematic curve – hardness curve – tolerances – sequence – file – scale – histogram |
Output interfaces | USB – Ethernet |
Reading | Direkt reading of case depth on touch-screen display 8.4” – Schematic curve with case depth value at the predetermined hardness – Stats and overlay of curves |
Languages | Italian, English, German, Other languages on request |
Load application method | Automatic with predetermined load value according to case depth |
Weight | approx. 440 kg |
Working principle | Evaluation of the load-penetration curve and tested algorythm |
Standard scales | Reference scale HV |
Load time | Test cycle depending on applied load |
Normal lagervare, forbehold for mellemsalg